  • DEUTSCH, seite 22
Mechanical alkaline cleaning and thermal disinfection
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection 90/194
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water
quality at least)
Place the product in a tray that is suitable for
cleaning (avoiding rinsing blind spots)
Connect components with lumens and channels
directly to the rinsing port of the injector carriage.
Keep working ends open for cleaning.
Check visible surfaces for residues after mechani-
cal cleaning/disinfecting.
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
5
-
-

Inspection, maintenance and checks

Water
Chemicals
quality
D–W
-
FD–W
Dr. Weigert neodisher® SeptoClean
1 %* working solution
FD–W
-
FD–W
-
-
In accordance with the program
for the cleaning and disinfecting
machine
Damage (metal seizure/fric-
tion corrosion) to the product
caused by insufficient lubrica-
CAUTION
tion!
Prior to function checks,
lubricate moving parts (e.g.
joints, pusher components
and threaded rods) with
maintenance oil suitable
for the respective steriliza-
tion process (e.g. for steam
sterilization: Aesculap
STERILIT® I oil spray JG600
or STERILIT® I drip lubrica-
tor JG598).
Allow the product to cool down to room tempera-
ture.
17