Mechanical alkaline cleaning and thermal disinfection
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection 90/194
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water
quality at least)
►
Place the product in a tray that is suitable for
cleaning (avoiding rinsing blind spots)
►
Connect components with lumens and channels
directly to the rinsing port of the injector carriage.
►
Keep working ends open for cleaning.
►
Check visible surfaces for residues after mechani-
cal cleaning/disinfecting.
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
5
-
-
Inspection, maintenance and checks
►
Water
Chemicals
quality
D–W
-
FD–W
Dr. Weigert neodisher® SeptoClean
1 %* working solution
FD–W
-
FD–W
-
-
In accordance with the program
for the cleaning and disinfecting
machine
Damage (metal seizure/fric-
tion corrosion) to the product
caused by insufficient lubrica-
CAUTION
tion!
►
Prior to function checks,
lubricate moving parts (e.g.
joints, pusher components
and threaded rods) with
maintenance oil suitable
for the respective steriliza-
tion process (e.g. for steam
sterilization: Aesculap
STERILIT® I oil spray JG600
or STERILIT® I drip lubrica-
tor JG598).
Allow the product to cool down to room tempera-
ture.
17