Safe handling and preparation
CAUTION
Federal law restricts this device to sale by, or on
order of a physician!
Risk of injury caused by incor-
rect operation of the product!
WARNING
Note
A specially designed Aesculap endoscope is offered for
the MINOP InVent trocar system. For reasons of com-
patibility, the MINOP InVent trocars may only be used
with this special Aesculap endoscope.
Ensure that the product and its accessories are
operated and used only by persons with the requi-
site training, knowledge, or experience.
Read, follow, and keep the instructions for use.
Use the product only in accordance with its
intended use, see Intended use.
Remove the transport packaging and clean the new
product, either manually or mechanically, prior to
its initial sterilization.
Store any new or unused products in a dry, clean,
and safe place.
Prior to each use, inspect the product for loose,
bent, broken, cracked, worn, or fractured compo-
nents.
Do not use the product if it is damaged or defec-
tive. Set aside the product if it is damaged.
Replace any damaged components immediately
with original spare parts.
Perform a function check of the irrigation channel
before each use.
Attend appropriate prod-
uct training before using
the product.
For information about
product training, please
contact your national
B. Braun/Aesculap agency.
Perform a function check of the endoscope's stop
mechanism before each use.
To avoid burns when using the MINOP InVent tro-
car in combination with HF electrodes, ensure that
HF current is activated only under visual control.
When inserting the trocar into the brain/the ven-
tricle, close the working channel with the obtura-
tor provided.

Safe operation

Risk of injury and/or malfunc-
tion!
WARNING
Risk of injury when using the
product beyond the field of
view!
WARNING
Risk of injury caused by pro-
truding 30° endoscope in
locked position!
WARNING
Risk of burns due to high tem-
perature of the instrument-
side end of the optical cable of
WARNING
a light source!
Always carry out a func-
tion check prior to using
the product.
Apply the product only
under visual control.
Insert the endoscope only
when the trocar is in its
final position.
Apply proper care when
operating the light source.