The blood and fluid pathways of the tubing set are sterilized with ethylene oxide and are
3.
nonpyrogenic. Under any of the following conditions, DO NOT use the set:
The end caps are not in place
Severe kinks in the tubing are apparent
The tubing set is incorrectly assembled
The tubing set is damaged
The clamps are closed on the pre-attached filter tubing assembly
When loading the channel and the LRS chamber into the centrifuge filler, be careful not to
4.
stretch the tubes, as tubing leaks could develop.
Be careful to press on tubing only with your fingers when loading the channel. To avoid
5.
accidental puncture, never use any sharp object.
When using the Trima Accel Automated Blood Collection System, Version 5, do not select a
6.
procedure configured with an RBC product when using a disposable tubing set that does not
contain an RBC bag. If a procedure with an RBC product is selected in error, a "centrifuge
pressure high" alert will be displayed when RBC collection begins. Select for rinseback and
continue. Any RBCs in the system will be returned to the donor.
If the tubing set remains functionally closed, the Platelet products can be stored for up to five
7.
days. Country specific regulations may allow for storage up to seven days.
Procedures for Use
Refer to the Trima Accel Automated Blood Collection system operator's manual for specifics
regarding the following procedures:
Setting up the disposable tubing set
Performing a collection procedure
Removing products and the disposable tubing set
Supplemental Instructions for Connecting Anticoagulant (AC)
When you are prompted to connect AC, connect the luer on the AC line (orange) to the AC bag.
Break the frangible connector on the AC bag to allow the AC to flow. Squeeze the drip chamber in
order to fill it approximately halfway. Follow the system prompts to load the AC line into the AC
sensor, and continue according to the Trima Accel system operator's manual.
Note: When you break frangible connectors, bend them in both directions to ensure that you break them
completely. Failure to do so may result in restricted flow.