Intended Use
The Trima Accel
LRS
®
Blood Collection System to collect platelets and plasma.
Please refer to the Trima Accel
complete listing of warnings, cautions, adverse effects and operating instructions.
Warnings
1.
If the integrity of the tubing set is compromised for any reason, the tubing set is no longer
functionally closed, and the product should not be stored beyond 24 hours.
2.
Check that all pump headers have been properly loaded before beginning a procedure.
Visually inspect each pump and carefully run your finger around the edge of each pump.
3.
This product contains phthalates (DEHP). Whole blood donors are not exposed to DEHP; the
potential health risk to apheresis donors is low, because the time-averaged DEHP dose
exposure is very low. Patient groups that include pregnant or nursing women and children are
considered to be the most at risk to potential harmful effects of exposure to DEHP. However,
regulatory bodies have noted that the benefit of doing a needed procedure is far greater than
the risk associated with exposure to DEHP. It is the responsibility of the treating physician to
balance this risk for the patient.
4.
Do Not Reuse/Not for Reuse: Terumo BCT, Inc. products bearing the "Do Not Reuse"
symbol are intended for single use only and are not intended to be reused or re‑sterilized in
any manner. Terumo BCT cannot ensure the functionality or sterility of the product if it is
reused or re‑sterilized.
Reuse of a single‑use product could result in:
Product performance issues due to a loss of product integrity, including but not limited
to the following:
Fluid leaks
Parts that are warped or deformed
Plastics that are brittle and discolored
Filters that have reduced filtration capabilities
Exposure to excessive ethylene oxide (EtO) residuals
Viral infections such as hepatitis or human immunodeficiency virus (HIV)
Bacterial infections
Cross‑contamination
Any of these risks could result in serious injury or death. These risks are shared by product
users, donors, patients, and recipients of end products of the device.
5.
The blood and fluid pathways of the tubing set are sterilized with ethylene oxide and are
nonpyrogenic. The packaging cover of the tubing set serves to protect the set and keep
contents in place during shipping. Under any of the following conditions, DO NOT use the
set:
The end caps are not in place
The needle cap plug is not in place
Platelet, Plasma Set is intended for use with the Trima Accel Automated
®
®
Automated Blood Collection System operator's manual for a
2