Severe kinks in the tubing are apparent
•
The tubing set is incorrectly assembled
•
The tubing set is damaged
•
The clamps are closed on the pre-attached filter tubing assembly
Cautions
1.
Each operator should be thoroughly familiar with the Trima Accel Automated Blood
Collection system operator's manual. All procedures should be performed by qualified
medical personnel under the supervision of a physician.
2.
Use aseptic technique throughout all procedures to ensure donor safety and product quality.
3.
When loading the channel and the LRS chamber into the centrifuge filler, be careful not to
stretch the tubes, as tubing leaks could develop.
4.
Be careful to press on tubing only with your fingers when loading the channel. To avoid
accidental puncture, never use any sharp object.
5.
When using the Trima Accel system Version 5, do not select a procedure configured with an
RBC product when using a disposable tubing set that does not contain an RBC bag. If a
procedure with an RBC product is selected in error, a "centrifuge pressure high" alert will be
displayed when RBC collection begins. Select for rinseback and continue. Any RBCs in the
system will be returned to the donor.
Procedures for Use
Refer to the Trima Accel Automated Blood Collection system operator's manual for specifics
regarding the following procedures:
•
Setting up the disposable tubing set
•
Performing a collection procedure
•
Removing products and the disposable tubing set
Storage Conditions
•
Shipping temperature range: –29 °C to 60 °C
•
Long-term storage temperature range: 0 °C to 60 °C
•
The disposable tubing sets are acceptable for use up to 24 hours after the set packaging is
opened (i.e., the Tyvek
Symbols and Certifications
Symbol
is removed, allowing access to the end caps).
®
Indicates the product quantity when the quantity is placed in the square.
Indicates the manufacturer of the product when accompanied by the name of the
manufacturer.