Ultrasonic Unit SonicWeld Rx®
Instructions for Use
Guidance and manufacturer's declaration – Electromagnetic interference emission
The SonicWeld Rx® is intended for use in an electromagnetic environment as specified below. The user should assure that it is
used in such an environment.
Immunity test
Electrostatic discharge
(ESD) according to
IEC 61000-4-2
Fast electrical
transients/bursts
according to
IEC 61000-4-4
Surge according to
IEC 61000-4-5
Voltage dips, short power
interruptions and supply
voltage fluctuations
according to
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
according to
IEC 61000-4-8
Note
Revision 04
IEC 60601 test level
±8 kV contact discharge
±15 kV air discharge
±2 kV for mains power
lines
±1 kV for input and
output lines
±0.5 kV differential-mode
voltage
±1 kV differential-mode
voltage
±2 kV common-mode
voltage
0% U
for ½ period
T
at 0 – 315 degrees
0% U
for 1 period at
T
0 degrees
70% U
for 25 periods at
T
0 degrees
0% U
for 250 periods
T
30 A/m
U
is the a. c. mains voltage prior to application of the test level.
T
Compliance level
±8 kV contact discharge
±15 kV air discharge
±2 kV for mains power
lines
±1 kV for input and
output lines
±0.5 kV differential-mode
voltage
±1 kV differential-mode
voltage
±2 kV common-mode
voltage
0% U
for ½ period
T
at 0 – 315 degrees
0% U
for 1 period at
T
0 degrees
70% U
for 25 periods at
T
0 degrees
0% U
for 250 periods
T
30 A/m
Electromagnetic
environment – Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user requires continued
operation during mains power
interruptions, it is recommended to
power the device from an
uninterruptible power supply.
Power-frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.