Ultrasonic Unit SonicWeld Rx®
Instructions for Use
4.3.5

Automatic Functional Test

Danger of burns!
If the sonotrode is allowed to make skin contact during the functional test, this can cause burns.
• Avoid contact with skin, eyes, etc.!
Self-test: Upon turning on the unit with the On/Standby switch (4, front fold-out page), the handpieces (if
connected) are ready for a self-test. This is indicated for each handpiece by the flashing activation LED (9,
front fold-out page) and the following symbol flashing on the corresponding side of the display:
The self-test for this handpiece will be started upon operating the hand activation pushbutton (10,
front fold-out page) for the first time. Hold down the button until the display message changes from
Testing to Cooling, the activation LED stops flashing and the acoustic signal changes. Be sure to keep the tip
of the sonotrode out of contact with objects during this process. If the test is successful, the unit is
automatically set to working mode.
• Checkmark (√) displayed (checkmark right or left, depending on which handpiece has been tested):
Means "no malfunction detected".
• Generator-related error messages are displayed in text form, while handpiece states are indicated by
pictogram on the respective side (depending on handpiece), see section 6.1, "Troubleshooting",
page 77.