Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data - Kls Martin Group Sonicweld Rx Gebrauchsanweisung

Ultraschallgerät
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 11
Ultrasonic Unit SonicWeld Rx®
Instructions for Use

9 Technical Data

Designation
Mains fuse (2 pieces)
Power supply frequency
Power supply voltage
Power input
Maximum power
Operating frequency
Protection class
Degree of protection
Suitability for use in atmospheres enriched
with oxygen
Operating mode
Applied part
Dimensions
Length of handpiece cable
Length of long handpiece cable
Number of reprocessing cycles for the
handpiece
Environmental conditions for transport and
storage
Environmental conditions for operation
Revision 04
SonicWeld Rx®
T 2 AH, 250 V
50–60 Hz
100–240 V alternating voltage with protective ground contact
70 VA
17 W
30 kHz
I acc. to IEC 60601-1
IP42
The SonicWeld Rx® is not suitable for use in atmospheres enriched with
flammable mixtures of anesthetics and air, oxygen or nitrous oxide.
Continuous duty with intermittent loading
2 s operation
15 s break
5 repetitions
4 min. break
Handpiece with type BF sonotrode
Width
Height
Depth
2.95 m
6 m
250
Ambient temperature
Relative humidity
Atmospheric pressure
Ambient temperature
Relative humidity
Atmospheric pressure
Altitude
Conforms to 93/42/EEC
260 mm
255 mm
46 mm
−20°C to +45°C
(−4°F to +113°F)
10–100%
500–1,060 hPa
+10°C to +40°C
(+50°F to +104°F)
20–75%
700–1,060 hPa
max. 4,000 m
85