Herunterladen Inhalt Inhalt Diese Seite drucken

Putting The Unit Out Of Operation - Kls Martin Group Sonicweld Rx Gebrauchsanweisung

Ultraschallgerät
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 11
The smoothing sonotrodes can also be used for smoothing out sharp or projecting edges of the membrane
or foil.
• Place the tip of the smoothing sonotrode at the edge of the membrane or foil.
• Exert just a little pressure on the edge of the membrane or foil.
• Operate the activation pushbutton until the membrane liquefies visibly. As long as the pushbutton is
pressed down, the activation LED (9, front fold-out page) lights up on the handpiece and an acoustic
signal is heard in addition.
• Run the tip of the sonotrode very slowly along the edge of the membrane/foil during activation of the
system.
If activation is continued for 60 s, the handpiece will be switched off automatically and the activation
LED (9, front fold-out page) will go out.
4.5

Putting the Unit out of Operation

To put the unit fully out of operation:
• Operate the On/Standby switch (4, front fold-out page).
− This sets the unit to standby mode and the display (3, front fold-out page) turns dark.
• Disconnect all system components (sonotrode from handpiece, handpiece with connecting cable from
ultrasonic unit and mains cable from ultrasonic unit).
For all-pole disconnection of the unit from the power supply, pull the mains plug.