Replacements

6.2.1

Replacing the Sonotrode

• Use the wrench (12, front fold-out page) to remove the sonotrode (11, front fold-out page) from the
handpiece by counterclockwise rotation; subsequently tighten the new sonotrode with the wrench
(clockwise rotation).
6.2.2

Replacing the Mains Fuse

• Set SonicWeld Rx® to standby mode by pressing the On/Standby switch.
− The unit can now no longer be activated and the display turns dark upon entering standby mode.
• Remove mains cable from mains socket-outlet.
• Pull out the fuse holder (6, front fold-out page).
• Remove the defective mains fuse from the fuse holder.
• Insert new mains fuse into the fuse holder, see section 8 "Accessories & Spare Parts", page 84, into the
fuse holder.
• Push fuse holder back into the unit, making sure that it locks in place.
• Check the unit for proper functioning.
6.2.3

Replacing the Activation LED

The activation LED (9, front fold-out page) cannot be replaced by the user. However, a defective LED has no
adverse effect on the proper functioning of the SonicWeld Rx® system as a whole.
6.2.4

Replacing the Mains Cable

• Set SonicWeld Rx® to standby mode by pressing the On/Standby switch. The unit can now no longer be
activated. The display turns dark upon entering standby mode.
• Unplug the mains cable first from the mains socket-outlet and then from the IEC inlet connector (5,
front fold-out page).
• Use new country-specific mains cable (see section 8 "Accessories & Spare Parts", page 84) and connect
it first to the unit (5, front fold-out page) and then to a grounding outlet.
• Check the unit for proper functioning.