• Once the SonicPin has solidified, remove the sonotrode tip with a twist.
Resorb x® or Resorb xG mesh or foils can also be fixed as follows:
• First, implant a sufficient number of SonicPins in the bone as described above, using the standard
sonotrodes (11, 13, front fold-out page) for pin insertion.
• Put the Resorb x® or Resorb xG membrane or foil in place over the inserted pins.
• Use the smoothing sonotrode (14, 15, front fold-out page) or the angled working end of the standard
sonotrode (13, front fold-out page) to weld the Resorb x® or Resorb xG membrane or foil in place on the
inserted pins.
− Place the tip of the smoothing sonotrode onto the membrane/foil at the point of contact between the
membrane/foil and the SonicPin underneath.
− Exert just a little pressure on the membrane/foil.
− Operate the activation pushbutton until the membrane liquefies visibly. As long as the pushbutton is pressed
down, the blue activation LED (9, front fold-out page) lights up on the handpiece and an alert is sounded in
addition.
− Do not move the tip of the sonotrode for at least 3 s after operating the pushbutton.
− Once the membrane/foil has solidified again, withdraw the tip of the sonotrode.
• Repeat this procedure to fix the Resorb x® or Resorb xG membrane or foil to all the remaining SonicPins.