Ultrasonic Unit SonicWeld Rx®
Instructions for Use
6.1.3

Potential Malfunctions

Trouble
"Screeching" noises heard
during the self-test or during
use of the SonicWeld Rx®
system.
Sonotrode tip will not detach
from SonicPin.
SonicPin fails to attach to
sonotrode tip.
Sonotrode tip sticks to
SonicPin.
Handpiece indicator does not
light up during activation.
Display dark.
Revision 04
Cause
Residual moisture or
condensate on the contact
points.
Bent or rammed
sonotrode tip.
Ultrasound applied too
long.
Bent sonotrode tip.
LED life exceeded.
No power supply.
Mains fuse defective.
Unit not switched on.
Identification
Clearly audible.
Visual check.
Pin is generally
deformed, sonotrode
cannot be removed
residue-free.
Visual check.
Visual check.
Visual check.
Electrical test.
Visual check.
Remedy
This does not compromise the
proper functioning of the
SonicWeld Rx® system. Noise will
subside automatically during
normal use of the device. It may be
helpful to use longer drying times
after sterilization.
Replace sonotrode tip.
Observe required cooling-down
time. Rotate the sonotrode axially
(twist) to break the connection,
see section 4.4, page 64.
Replace sonotrode tip.
Replace the handpiece. A defective
indicator on the handpiece has no
adverse effect on ultrasound
application.
Connect the system to the power
supply and operate the
On/Standby switch.
Replace mains fuse.
Press On/Standby switch until
display lights up.