Turn the product 1 with the battery shaft 4 upwards.
Press both safety catches 9 on the lid 8 at the same time and remove
the lid 8.
Tightly grasp the battery shaft 4 at the lower end.
Tap the battery shaft 4 on the palm of the hand until the battery 7
slides out of the battery shaft 4 and can be removed.
6.1.5

Protection against inadvertent activation

To prevent inadvertent activation of the product during tool change, the
trigger can be locked.
To lock trigger:
Twist the safety catch 3 to position OFF.
The trigger 2 is blocked and the product 1 cannot be operated.
To unlock trigger:
Twist the safety catch 3 to position ON.
The trigger 2 is unlocked and the product 1 can be operated.
6.1.6

Attaching and detaching the saw blade

Risk of injury when attaching/removing sawblades
in the ON position through inadvertent activation
of the product!
WARNING
Only attach/remove tools in the OFF position.
Coupling the saw blade
Secure product 1 against inadvertent activation with the safety
catch 3.
Insert the TPLO saw blade 11 all the way into the tool holder 14 with
the connection side in the desired position,
Secure the TPLO saw blade 11 with the clamp screw 12 using the pin
spanner 13.
Uncoupling the saw blade
Secure product 1 against inadvertent activation with the safety
catch 3.
Loosen the clamp screw 12 with the pin spanner 13.
Pull out TPLO saw blade 1 from the tool holder 14.
6.2

Function checks

The function checks must be carried out prior to each use and after each
intraoperative battery change.
Make certain there is a battery 7 inserted.
Ensure that the lid 9 has completely clicked into place.
Pull on the TPLO saw blade 11 to check that the TPLO saw blade 11 is
securely coupled.
Check that the cutting edges of the TPLO saw blade 11 do not exhibit
any mechanical damage.
Release product 1 for use (ON position).
Briefly operate product 1 at maximum oscillation frequency.
Do not use the product 1 if it is damaged or defective.
Set aside the product 1 if it is damaged.
6.3

Safe operation

Coagulation of patient tissue or risk of burns for
patients and user through hot product!
Cool the tool during operation.
WARNING
Keep product/tool out of the reach of patients.
Let the product/tool cool down.
Use a cloth to protect against burns when
changing the tool.
Risk of infection or injury through aerosol forma-
tion or particles ejected from the tool!
Use suitable protection (such as waterproof pro-
WARNING
tective clothing, face mask, safety gases, suc-
tion).
Risk of injury and/or malfunction!
Always carry out a function check prior to using
WARNING
the product.
Risk of injury when using the product beyond the
field of view!
WARNING
Apply the product only under visual control.
Risk of injury and damage to the tool/system!
The tool may get caught in drapes (such as tex-
WARNING
tiles).
Do not let the tool come into contact with
drapes (such as textiles) during operation.