Before attempting the insertion, ensure that you are familiar with
the potential complications and their emergency treatment should
any of them occur.
WARNINGS:
•
In the rare event that a hub or connector separates from any
component during insertion or use, take all necessary steps and
precautions to prevent blood loss or air embolism and remove
catheter.
•
Do not advance the guidewire or catheter if unusual resistance is
encountered.
•
Do not insert or withdraw the guidewire forcibly from any
component. The wire may break or unravel. If the guidewire
becomes damaged, the introducer needle or Vascu-Sheath
introducer and guidewire must be removed together.
•
Federal Law (USA) restricts the device to sale by or on the order of a
physician.
•
This catheter is for Single Use Only.
•
Do not resterilize the catheter or accessories by any method.
•
Re-use may lead to infection or illness/injury.
•
The manufacturer shall not be liable for any damages caused by re-
use or resterilization of this catheter or accessories.
•
Contents sterile and non-pyrogenic in unopened, undamaged
package. STERILIZED BY ETHYLENE OXIDE
•
Do not use catheter or accessories if package is opened or damaged.
•
Do not use catheter or accessories if any sign of product damage is
visible.
CATHETER PRECAUTIONS:
•
Do not use sharp instruments near the extension tubing or catheter
lumen.
•
Do not use scissors to remove dressing.
•
Catheter will be damaged if clamps other than what is provided with
this kit are used.
•
Clamping of the tubing repeatedly in the same location may weaken
tubing. Avoid clamping near the luers and hub of the catheter.
•
Examine catheter lumen and extensions before and after each
treatment for damage.
•
To prevent accidents, assure the security of all caps and bloodline
connections prior to and between treatments.
•
Use only Luer Lock (threaded) Connectors with this catheter.
•
Repeated over tightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector failure.
INSERTION SITES:
Warning:
Physician discretion is strongly advised when inserting this
catheter in patients who are unable to take or hold a deep breath.
•
The patient should be in a modified Trendelenburg position, with
the upper chest exposed and the head turned slightly to the side
opposite the insertion area. A small rolled towel may be inserted
between the shoulder blades to facilitate the extension of the chest
area.