TROUBLESHOOTING
36
In-Test Error Dialogs
Cartridge/Carrier Could Not Be
Correctly Found - Scale
Error: 36
The instrument fiducial features on the
cartridge/strip carrier found but the
scale is out of range.
Cartridge/Carrier Could Not Be
Correctly Found - Position
Error: 37
The instrument fiducial features on
the cartridge/strip carrier found but
position is out of range.
Failed to locate the strip cartridge/
carrier
Error: 38
The instrument fiducial features on the
cartridge/strip could not be found. The
test was unable to proceed.
Cartridge/Carrier Could Not Be
Correctly Found - Rotation
The analysis software was able to
Error: 41
locate the cartridge/strip carrier;
however the rotation angle found was
outside of the acceptable limits.
Could not locate the control line,
Multiple Candidates
The instrument was unable to
determine the location of the control
Error: 43
line with sufficient confidence. The
algorithm found more than one control
line.
This error may occur when no control
line is present on the strip.
Strip Analyser Could Not Locate
Control Line, No Line Found
Error: 44
The instrument has found no control
lines within the control line search
region.
Check that the correct cartridge/strip carrier is being used
with the instrument.
Check that the cartridge/strip carrier was inserted correctly
into the instrument.
Check for contaminants.
Check that the selected test matches the cartridge/carrier
used.
Attempt to rerun test using a new cartridge/strip carrier.
If the error continues, contact the supplier.
Check that the correct cartridge/strip carrier is being used
with the instrument.
Check that cartridge/strip carrier was inserted correctly
into the instrument.
Check for contaminants.
Check that the selected test matches the cartridge/carrier
used.
Attempt to rerun test using a new cartridge/strip carrier.
If the error continues, contact the supplier.
Check that the correct cartridge/strip carrier is being used
with the instrument.
Check that cartridge/strip carrier was inserted correctly
into the instrument.
Check for contaminants.
Check that the selected test matches the cartridge/carrier
used.
Attempt to rerun test using a new cartridge/strip carrier.
If the error continues, contact the supplier.
Check that the correct cartridge/strip carrier is being used
with the instrument.
Check that the cartridge/strip carrier was inserted correctly
into the instrument.
Check for contaminants.
Check that the selected test matches the cartridge/carrier
used.
Attempt to rerun test using a new cartridge/strip carrier.
If the error continues, contact the supplier.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Check for control line being present.
Check the position of the lines on the strip to ensure that it
is within manufacturing specifications/tolerances.
Check that a test line is not within the search region for the
control line.
Check for contaminants on the strip.
Repeat testing with new dipstick.
If the error continues, contact the supplier.
The Supplier will review the test types package.
Check that the strip is correctly inserted in the cartridge/
strip carrier.
Check for the control line.
Check the position of the lines on the strip to ensure that it
is within manufacturing specifications/tolerances.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package.