TROUBLESHOOTING
38
In-Test Error Dialogs
Could Not Complete Normalisation,
Line Separation Below Limit
Error: 59
The normalisation failed to complete.
The Normalisation lines appear to be
too close together.
Could Not Make Decision, Unknown
Exception
Error: 64
The decision module in the test type
encounters an error.
Could Not Make Decision, Audio Tone
Not Set
This is most likely due to an error in
Error: 65
the logic of the decision algorithm
definition in the test package. The
Audio Tone is not being set by the
algorithm and is required.
Could Not Make Decision, Detailed
Message Not Set
This is most likely due to an error in
Error: 66
the logic of the decision algorithm
definition in the test package.
The required 'Detailed Message' Fields
are not set by the algorithm.
Could Not Make Decision, Icon Not Set
This is most likely due to an error in
the logic of the decision algorithm
Error: 67
definition in the test package.
The required 'Icon' fields are not set by
the algorithm. (RESULT_DECISION_
ICON_X)
Could Not Make Decision, Message
Not Set
This is most likely due to an error in
the logic of the decision algorithm
Error: 68
definition in the test package.
The required 'Message' fields are not
set by the algorithm.
(RESULT_DECISION_MESSAGE_X)
Check normalisation strip. Ensure that the correct strip is
being used.
Ensure strip is located at the mid-point of the carrier/
correctly mounted in the cartridge and that lines are clearly
visible.
Re-attempt normalisation using a different cartridge/
carrier.
Contact Manufacturer if error persists.
This is most likely due to an error in the logic of the decision
algorithm definition in the test package.
Other Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the audio tone is set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the Detailed Message fields are set correctly by the
algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the result Icons are set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the result messages are set correctly by the algorithm.