In-Test Error Dialogs
Could Not Make Decision, Title Not Set
This is most likely due to an error in
the logic of the decision algorithm
Error: 69
definition in the test package.
The required 'Title' fields are not set by
the algorithm
(RESULT_DECISION_TITLE_X)
Could Not Make Decision, Set UI Type
Not In Allowed List
This is most likely due to an error in
Error: 70
the logic of the decision algorithm
definition in the test package.
The required 'Type' field is not set to a
valid value by the algorithm.
Could Not Make Decision, UI Type
Not Set
This is most likely due to an error in
Error: 71
the logic of the decision algorithm
definition in the test package.
The required 'Type' field is not set by
the algorithm.
Could Not Make Decision, Unknown
Type
This is most likely due to an error in
Error: 72
the logic of the decision algorithm
definition in the test package.
The required 'Type' field is set to an
invalid type for the application.
Could Not Make Decision, Valid Flag
Not Set
This is most likely due to an error in
Error: 73
the logic of the decision algorithm
definition in the test package.
The required 'Valid' field is not set to
true by the algorithm.
Could Not Make Decision, Ratio Title
Not Set
This is most likely due to an error in
Error: 74
the logic of the decision algorithm
definition in the test package.
The required 'Ratio Title' field is not set
by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the result 'Title' fields are set correctly by the algorithm
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the UI type for the result is set correctly by the
algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the UI Type is set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the UI Type is set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the UI Type is set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the required number of ratio output fields are defined
correctly.
TROUBLESHOOTING
39