Errors Dialogs
Starting remote mode failed!
0519
Please try again.
Passwords do not match!
0522
Please re-enter password.
Import of user list failed!
0523
Please try again.
Export of user list failed!
0524
Please try again.
Import failed!
0526
Multiple test type packages detected.
Please check USB key and try again.
No test type package found!
0527
Please check USB key contents and try
again.
A maximum of 50 test types can be
0528
imported onto the instrument!
Please edit the test types package
Test types package import failed!
0529
Please try again.
SD card not found!
0530
The instrument will reboot.
Please refer to user manual.
Control status has failed!
0532
Please update the control
status.
Exposure calibration failed!
0533
Please try again.
Normalisation failed!
0534
Please try again.
Instrument check has failed!
0535
Please run a new instrument check.
If the user attempts to establish a remote connection with
the Actim 1ngeni instrument connected via network to the
kinetic software and the attempt fails, it is suggested that
repeated attempts are made once the user establishes
that the network and physical hardware setup is configured
correctly.
The password entered does not match the password saved
on the instrument. Please try to attempt entering the
password again. If you have forgotten your password, please
contact the supplier.
The import of the user list failed. Please check that the USB
key is correctly connected to the instrument and re-attempt
import.
The export of the user list failed. Please check that the USB
key is correctly connected to the instrument and re-attempt
export.
The instrument has detected that there is more than one
test types package available for import on the attached USB
key. Remove one of the test types packages from the USB
key then re-attempt import.
If no Test types are loaded onto the instrument then a Test
Type Package must be imported onto the instrument from
an attached USB Flash memory key.
The test types package the user is attempting to import is
too large. Please contact the supplier.
The supplier must ensure that test types package provided
to the end user contains less than 50 tests.
The instrument was unable to import the Test Types from
an attached USB Flash memory key. Please ensure that the
file is placed correctly in the main file directory and the file
name correct. Ensure that only a single test type package is
located on the USB key.
The Actim 1ngeni instrument cannot find the external SD
memory card.
The instrument will not boot up.
Arrange for the return of the instrument.
The control status for the selected test type has failed.
Please run a new control test for that test type to update the
control status to pass.
The attempt made to calibrate the Actim 1ngeni instrument
has failed. Please re-attempt calibration. If this does not
correct the issue then contact the manufacturer.
The attempt made to normalise the Actim 1ngeni
instrument has failed. Please re-attempt normalisation. If
the problem persists, replace the normalisation strip. If this
does not correct the issue then contact the manufacturer.
The user may not* be able to run a test until the instrument
check test has passed.
Run a new instrument check to update the instrument
check status to Pass.
*Instrument check functionality is configurable and the
settings may vary.
TROUBLESHOOTING
43