In-Test Error Dialogs
CODE
DESCRIPTION
Could Not Identify Cartridge
This can occur only if the overall image
Error: 2
average grey level is below a defined
limit.
Could Not Locate Validation Marks
The instrument could not locate the
internal image validation targets.
Error: 3
This can occur only if the test has
been configured to 'Run Validate Mark
Check' using kinetic and the valuator
marks cannot be found in the image.
Could Not Locate Strip
The instrument could not locate the
Error: 5
cartridge/strip carrier in the acquired
image.
Validation Mark Positions Outside Of
Calibrated Limits
Error: 25
The instrument internal image
validation targets were found, but were
not in an acceptable position.
Validation Mark Value Out Of Absolute
Limits
The instrument internal validation
Error: 32
mark targets are out of the acceptable
limits. This can only occur if the test
has been configured with the 'Run
Validation Check' option selected.
Validation Mark Value Out Of Calibrated
Limits
The instrument internal validation
Error: 33
mark targets are out of the acceptable
limits. This can only occur if the test
has been configured with the 'Run
Validation Check' option selected.
ACTION
Possible cause: the strip is not present in the strip carrier.
Check that the strip is in the cartridge/strip carrier.
Check for contaminants.
Attempt to rerun test
Possible cause: the camera or LED's have failed and the
image is black
Run a self test on the instrument.
If the self test fails, contact the supplier
Possible cause: exposure calibration is not correct – re-
execute the exposure calibration.
Run a self test on the instrument.
If the instrument needs to be calibrated then contact the
supplier.
Possible causes: Camera failure or Incorrect Validation
Mark Strips
Run a self test on the instrument.
If the self test fails, contact the supplier
"Acquire image" of the instrument internals and check the
condition of the validation mark strips.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Change it to check for contaminants on the strip carrier/
cartridge fiducial marks.
Attempt to rerun test
Try another strip carrier/cartridge.
If the error continues, contact the supplier.
"Acquire image" of the instrument internals and check the
image quality is adequate.
Check that no foreign bodies are inside of the instrument
Run Self Test
If the error continues, contact the supplier.
Supplier to acquire the image and ensure that the internal
validation marks are clearly visible and not obscured by
debris or other materials.
If it continues to occur, contact the manufacturer
Replace validation mark targets
Check that no foreign bodies are inside of the instrument
Run Self Test
If the error continues, contact the supplier.
Supplier to acquire image and ensure that the internal
validation marks are clearly visible and not obscured by
debris or other materials.
If it continues to occur contact the manufacturer
Replace validation mark targets
Check that no foreign objects are inside of the instrument
Run Self Test
If the error continues, contact the supplier.
Supplier to acquire image and ensure that the internal
validation marks are clearly visible and not obscured by
debris or other materials.
Contact manufacturer.
TROUBLESHOOTING
35