Errors Dialogs
Results data has been corrupted!
0563
Please refer to user manual.
The test has timed out!
0564
Please discard the test.
Password not recognised!
0565
Please try again.
IP address {0} is invalid!
0566
Please enter a valid IP address.
Subnet mask {0} is invalid!
0567
Please enter a valid subnet mask.
Printer not found!
0569
Please check printer connection and try
again.
Instrument is not normalised!
0570
Testing is locked out.
Please refer to user manual.
Maximum of 99 users reached!
0572
Please delete an existing user ID before
adding a new user.
There are no associated control tests for
0582
this test type!
Please discard the test.
Information Dialog
0256
User deletion completed successfully!
0258
{0} users imported successfully!
Exported {0} users successfully to USB
0259
instrument!
No user ID list found!
0260
Please check USB key and try again.
No user ID list found on instrument!
0261
Please enter users.
Test types imported successfully!
0262
Instrument shall reboot.
A results corruption has occurred, please attempt to export
the results. Please contact your supplier.
The next test step was not completed in the allotted time.
The test has timed out. Please discard the test.
The password entered is not recognised by the instrument,
please enter the correct password.
If you have forgotten the admin password, please contact
your supplier.
An incorrect or invalid IP address has been entered. Please
check what was entered then try again.
An incorrect or invalid subnet mask has been entered.
Please check what was entered then try again.
If the user attempts to print a test report before setting
up the instrument printer connections, this error will be
displayed.
a) Try connecting the USB report printer to the instrument.
If the printer is not found after connecting, try to reboot the
instrument.
b) Set up a network connected printer.
If the instrument is not successfully normalised prior to
testing, the test results may be incorrect. This instrument
requires normalisation to be complete before the Actim
1ngeni instrument can perform testing.
The instrument is notifying the user that the user list is full.
No more users can be added until some are cleared from
the instrument memory. Please delete a user to be able to
add a new user.
If the user attempts to run a control test for a test type that
does not have any associated control test types, then the
instrument shall notify the user. The user should dispose
of the test.
The delete user function has completed successfully.
User import function has completed successfully.
User export function has completed successfully.
If the user is attempting to import a user list and the instru-
ment cannot find the user list on the attached USB key then
the user should check the USB key contacts to ensure that
the file is in the correct location and in the correct format.
If there are no user ID entered on the instrument, the in-
strument shall inform the user. The admin user should add
users to the instrument then reattempt request.
The instrument has successfully imported Test Types. The
instrument informs the user that an instrument reboot is
required prior to testing with the newly imported Test Types.
TROUBLESHOOTING
45