In-Test Error Dialogs
Strip Analyser Could Not Locate Line,
Line Width Validation Failed
A line has been found, but the line
Error: 45
width score is outside of the acceptable
limit defined in the test type. Either the
test strip/cartridge is invalid, or the
test type is incorrectly defined.
Strip Analyser Could Not Locate Line,
Line Peak Validation Failed
A line has been found, but the peak
Error: 46
score is outside of the acceptable limit
defined in the test type. Either the test
strip/cartridge is invalid, or the test
type is incorrectly defined.
Strip Analyser Could Not Locate Line,
Line Area Validation Failed
A line has been found, but the area
Error: 47
score is outside of the acceptable limit
defined in the test type. Either the test
strip/cartridge is invalid, or the test
type is incorrectly defined.
Strip Analyser Could Not Locate Line,
Line Position Validation Failed
A line has been found, but the position
Error: 48
score is outside of the acceptable limit
defined in the test type. Either the test
strip/cartridge is invalid, or the test
type is incorrectly defined.
Focus Out Of Range
The focus of the camera is outside of
limits.
Error: 49
This error will only occur if the test has
the 'Run Focus Check' option enabled
in kinetic.
Could Not Complete Normalisation,
Could Not Locate Enough Lines
Error: 56
The normalisation failed to complete.
The instrument appears to have found
too few normalisation lines.
Could Not Complete Normalisation,
Located Too Many Lines
Error: 57
The normalisation failed to complete.
The instrument appears to have found
too many normalisation lines.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Check the test line.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Check the test line.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Check the test line.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package.
Check that strip is correctly inserted in the cartridge/strip
carrier.
Check the test line.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package.
Re-power the instrument.
Attempt to run a self test.
Supplier to acquire image and ensure that the internal
validation marks are clearly visible and not obscured by
debris or other materials.
Check normalisation strip. Ensure that the correct strip is
being used.
Ensure strip is located at the mid-point of the carrier/
correctly mounted in the cartridge and that lines are clearly
visible.
Re-attempt normalisation using a different cartridge/
carrier.
Contact Manufacturer if error persists.
Check normalisation strip. Ensure that the correct strip is
being used.
Ensure strip is located at the mid-point of the carrier/
correctly mounted in the cartridge and that lines are clearly
visible.
Re-attempt normalisation using a different cartridge/
carrier.
Contact Manufacturer if error persists.
TROUBLESHOOTING