cleaners and disinfectants or solutions containing iodine,
chlorine or certain metal salts. Also, in solutions with pH values
above 11, the anodisation layer may dissolve.
Manual Cleaning/Disinfection Instructions
1. Prepare enzymatic and cleaning agents at the use-dilution and
temperature recommended by the manufacturer. Fresh
solutions should be prepared when existing solutions become
grossly contaminated.
2. Place instruments in enzymatic solution until completely
submerged. Actuate all moveable parts to allow detergent to
contact all surfaces. Soak for a minimum of twenty (20)
minutes. Use a nylon soft bristled brush to gently scrub
instruments until all visible debris is removed. Pay special
attention to hard-to-reach areas. Pay special attention to any
cannulated instruments and clean with an appropriate bottle
brush. For exposed springs, coils, or flexible features: Flood
the crevices with copious amounts of cleaning solution to flush
out any soil. Scrub the surface with a scrub brush to remove all
visible soil from the surface and crevices. Bend the flexible
area and scrub the surface with a scrub brush. Rotate the part
while scrubbing to ensure that all crevices are cleaned.
3. Remove the instruments and rinse thoroughly under running
water for a minimum three (3) minutes. Pay special attention to
cannulations, and use a syringe to flush any hard to reach
areas.
4. Place the instruments, fully submerged, in an ultrasonic unit
with cleaning solution. Actuate all moveable parts to allow
detergent to contact all surfaces. Sonicate the instruments for a
minimum of ten (10) minutes.
PKGI-42-T Effective 08-2019
5. Remove the instruments and rinse in deionised water for a
minimum of three (3) minutes or until all signs of blood or soil
are absent in the rinse stream. Pay special attention to
cannulations, and use a syringe to flush any hard to reach
areas.
6. Inspect instruments under normal lighting for the removal of
visible soil.
7. If visible soil is seen, repeat the sonication and rinse steps
above.
8. Remove excess moisture from the instruments with a clean,
absorbent, nonshedding wipe.
Combination Manual/Automated Cleaning and Disinfecting
Instructions
1. Prepare enzymatic and cleaning agents at the use-dilution and
temperature recommended by the manufacturer. Fresh
solutions should be prepared when existing solutions become
grossly contaminated.
2. Place instruments in enzymatic solution until completely
submerged. Actuate all moveable parts to allow detergent to
contact all surfaces. Soak for a minimum of ten (10) minutes.
Use a nylon soft-bristled brush to gently scrub instruments until
all visible debris is removed. Pay special attention to hard-to-
reach areas. Pay special attention to any cannulated
instruments and clean with an appropriate bottle brush. Note:
Use of a sonicator will aid in thorough cleaning of instruments.
Using a syringe or water jet will improve flushing of difficult-to-
reach areas and any closely mated surface.
3. Remove instruments from enzyme solution and rinse in
deionised water for a minimum of one (1) minute.
English– EN/ PAGE 22