Herunterladen Inhalt Inhalt Diese Seite drucken

Table 7: Guidance And Manufacturer's Declaration - Hologic Fluentpro Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für FluentPro:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 610
Immunity test
IEC 60601-1-2 Table 11:
Enclosure Port Immunity to
Proximity Magnetic Fields
according to IEC 60601-1-2
Note *: UT is the AC mains voltage prior to application of the test level. Note 1: At 80 MHz and 800 MHz, the higher
frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection of structures, objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Fluent Pro Fluid Management System is used
exceeds the applicable compliance level above, the Fluent Pro Fluid Management System should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the changing orientation or the location of the Fluent Pro Fluid Management System.
b) Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 7: Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Frequency ranges and test conditions for RF wireless communications equipment
Frequency band (MHz)
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
82
Fluent Pro Fluid Management System
IEC 60601
Compliance level
test level
30 kHz, 8 A/m, CW
30 kHz, 8 A/m, CW
Modulation
134.2 kHz, 65 A/m,
134.2 kHz, 65 A/m,
2.1 kHz Pulse
Modulation
13.56 MHz,
7.5 A/m, 50 kHz
Pulse Modulation
Pulse Modulation
Test frequency (MHz)
380-390
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
Supplementary Information
Electromagnetic environment
guidance
Modulation
2.1 kHz Pulse
Modulation
13.56 MHz,
7.5 A/m, 50 kHz
Modulation
a
Pulse – 18 Hz
FM ±5 kHz deviation 1 kHz
sine
Pulse – 217 Hz
Pulse – 18 Hz
Pulse – 217 Hz
Pulse – 217 Hz
Pulse – 217 Hz
Compliance level (V/m)
27
28
9
28
28
28
9
English