Replacing the Tissue Trap
3.8.3
3.8.3.1
3.8.3.2
3.8.3.3
46
Fluent Pro Fluid Management System

Figure 55: Remove Full Waste Bag

Hang a new Waste Bag onto the Waste Bag Hooks.
Attach the Out-FloPak Waste Bag Connector onto the Waste Bag by rotating the connector
clockwise.

Figure 56: Install Waste Bag Connector

If 'Missing Waste Bag' error is displayed, ensure the Waste Bag is installed properly on the
Hooks, touch the 'CLEAR' icon and continue with the procedure.
If the tissue trap appears to be full, by expanding outside of the internal basket, the tissue
trap is near capacity and must be changed. Failure to do so could result in loss of tissue
sample. The System will not alert you if the tissue trap is near capacity or full, please
monitor tissue trap for capacity.
Touch the 'PAUSE' icon to pause the system.
Open the lid of the Tissue Trap Canister.