After Procedure

Disassembly and Disposal

4.1
WARNING!
Fluent Pro Procedure Kits are single use disposables. Do not re-use or reprocess any of the components
contained within the single-use Procedure Kit.
Note: Comply with your facility's disposal and hygiene rules when disposing of Fluent Pro Procedure Kits, the
Tissue Trap, fluid collected, and the Waste Bag.
Breakdown MyoSure Tissue Removal Device (TRD)
4.1.1
4.1.1.1
4.1.1.2
4.1.1.3
4.1.1.4

Collect Pathology

4.1.2
4.1.2.1
4.1.2.2
4.1.2.3
English
Fluent Pro Fluid Management System
Remove the TRD from the hysteroscope.
Disconnect the MyoSure TRD suction tube from the Out-FloPak tubing.
Disconnect the drive cable of the MyoSure TRD from the MyoSure TRD Panel.

Figure 58: Disconnecting the MyoSure TRD

Dispose of the TRD according to your facility's protocol.
Locate a specimen container (not supplied).
Open the lid of Tissue Trap Canister.
Remove the Tissue Trap from the Canister (leaving basket behind) and place it into the
specimen container. If additional tissue is to be collected, refer to Section 3.8.3 for Tissue Trap
replacement.
After Procedure
51