Table 5: Electromagnetic Emissions Guidance

Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Electromagnetic Immunity Guidance and Manufacturer's Declaration
6.4.2
The Fluent Pro Fluid Management System is intended for use in the electromagnetic environment
specified below. The customer or the user of the Fluent Pro Fluid Management System should ensure
that it is used in such an environment. Non-observance of the instructions listed in this manual can
lead to damage or malfunction of the system.

Table 6: Electromagnetic Immunity Guidance

Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transients /
bursts IEC 61000-4-4
Surge IEC 61000-4-5
80
Fluent Pro Fluid Management System
Compliance
Electromagnetic environment - guidance
Group 1
The Fluent Pro Fluid Management System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause interference in nearby electronic equipment.
Group 2
The Fluent Pro Fluid Management System must emit electro-magnetic
energy in order to perform its intended function. Nearby electronic
equipment may be affected.
Class A
The Fluent Pro Fluid Management System is suitable for use in all
establishments other than domestic and those directly connected to
Complies
the public low-volt- age power supply network that supplies buildings
used for domestic purposes.
Complies
Note: The EMISSIONS characteristics of this equipment make it
suitable for use in industrial areas and hospitals (CISPR 11 class A). If
it is used in a residential environment (for which CISPR 11 class B is
normally required) this equipment might not offer adequate protection
to radio-frequency communication services. The user might need to
take mitigation measures, such as relocating or re-orienting the
equipment.
IEC 60601
Compliance level
test level
± 8 kV contact
± 8 kV contact
± 2, ± 4, ± 8, ±
± 2, ± 4, ± 8, ±
15 kV air
15 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input /
± 1 kV for
output lines
input/output lines
± 1 kV line(s) to
± 1 kV line(s) to
line(s)
line(s)
± 2 kV line(s) to
± 2 kV line(s) to
earth
earth
Supplementary Information
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.