Fluent Pro Fluid Management System
Waste Bag. It fits securely into the yellow Fluent Pro Out-FloPak Receptacle on the front
right side of the fluid management system. These connections allow the transfer of fluid from
the hysteroscope Outflow Channel, the MyoSure tissue removal device (TRD), and the
Under-Buttocks Drape (UBD) Port to the Waste Bag.
Tissue Trap
The Tissue Trap is designed to capture resected tissue throughout the procedure to allow
the tissue to be sent to pathology. The Tissue Trap Canister contains the Tissue Trap that
captures the resected tissue. Please monitor the Tissue Trap so it does not overfill. For
details on handling the Tissue Trap, see Section 3.8.3.
Waste Bag
The non-sterile Waste Bag is designed to capture waste fluid from hysteroscopic
procedures. The Waste Bag hangs on the Waste Bag Hooks at the bottom of the Fluent Pro
Fluid Management System. Hang only a single Waste Bag on the Waste Bag Hooks at a
time. Hanging more than one bag may impact fluid deficit accuracy. The Waste Bag includes
an attached cap. For details on replacing the Waste Bag, see Section 3.8.2.
If an accurate manual deficit assessment is required, pour fluid into calibrated container. The
markings on the Waste Bag are not intended as a measuring device, are only for general
reference and are not intended as a specific volumetric measuring device.