2.7.6.11 Touch 'NEXT' on the Touchscreen.
2.7.6.12 The Initial Prime screen is then displayed.
Priming the system
2.7.7
2.7.7.1
2.7.7.2
2.7.7.3
2.7.7.4
English
Fluent Pro Fluid Management System
Figure 32: Initial Prime Screen
Begin each procedure with a new, full Fluid Bag. Priming the system runs the pump for
approximately one (1) minute to purge air from the tubing and calibrate the hysteroscope.
Before priming, make sure that the Fluent Pro In-FloPak and Fluent Pro Out-FloPak are
properly connected as described in Section 2.7: Setup System.
The system should be primed at the beginning of a case. Refer to Section 2.7.9 for instructions
on how to Reprime the system.
To prime the system, perform the following steps:
2.7.7.4.1
Unclamp Fluid Bag.
Figure 33: Unclamp Fluid Bag
Procedure Setup