Replacing Components

3.8
Depending on the procedure, you may need to replace disposable components. This Section provides
information and instructions for replacing the Fluid Bag, Waste Bag, and Tissue Trap while using the Fluent Pro
System. To alert you of changes, the Fluent Pro Fluid Management System Fluid Bag Hooks and Waste Bag
Hooks contain built-in scales, and the system will prompt you if you need to take action. The System will not alert
you if the tissue trap is near capacity or full, please monitor tissue trap for capacity.

Adding the Fluid Bag

3.8.1
3.8.1.1
3.8.1.2
3.8.1.3
3.8.1.4
3.8.1.5
3.8.1.6
English
Fluent Pro Fluid Management System
On the System Setup screen, touch the 'LAST PROCEDURE DATA' icon.
The following final data will be displayed from the last procedure:
Fluid Deficit
Total Fluid In
Cutting Time

Figure 52: Last Procedure Data Screen

Note: In the event of a power failure, refer to Section 5.2 for retrieval of last procedure data.
The system does not need to be paused in order to add or change a Fluid Bag, as long as a
second bag is hung and unclamped.
Replace the Fluid Bag when the Fluid Bag is almost empty or when alerted by the system.
If leaving an empty bag on Hooks, ensure that empty Fluid Bag is clamped.
Hang only a single Fluid Bag on the Fluid Bag Hooks at a time.
A minimum of one Fluid Bag should always be unclamped when the system is running.
Perform the following steps to replace the Fluid Bag during a procedure:
WARNING!
Do not place excessive force or weight on the Fluid Bag Hooks. Doing so may result in an
inaccurate fluid deficit value, causing risk to patient safety.
Only hang one (1) Fluid bag per supply hook. Hanging more than 1 fluid bag per supply
hook can affect the system's ability to recognize empty fluid bags during procedure.
3.8.1.6.1
One Fluid Bag on Fluid Bag Hooks
3.8.1.6.1.1
Hang a new Fluid Bag on the vacant hook.
During Procedure