6.1.2.12 If the value is within range press the 'NEXT' icon and complete the breakdown steps as
Technical Specifications
6.2
Table 4: Technical Specifications
Item
Model or type designation
Mains voltage range [V]
Supply Frequency Range [Hz]
Fuse designation
Mode of Operation
Power Consumption
Upper Voltage Range
Normal Operation
Peak
Lower Voltage Range
Normal Operation
Peak
Protection class (I, II, III)
Application part type (B, BF, CF)
Defibrillator protected (yes, no)
Protection type (IP code)
Conformity with the following
standards
Operating conditions
Storage and transportation
conditions
Use possible with flammable
anesthetic gases
Maximum sound level
Adjustable Values
Pressure range (mmHg)
Deficit Limit (mL)
78
Fluent Pro Fluid Management System
described. Press 'DONE' when complete.
Specification
FLT-200
100-240 VAC
50-60Hz
T5AH, 250 V fuses
Non-continuous (Max Duty: 60 minutes ON, 30 minutes IDLE)
350W
240 VAC
264 VAC
120 VAC
90 VAC
I
BF
No
IP21
IEC 60601-1:2005 + A1:2012 + A2:2020
IEC 60601-1-6:2010 + A1:2013 + A2:2020
IEC 62366-1:2015 + A1:2020
IEC 62304:2006 +A1:2015
IEC 60601-1-2:2014 + A1:2020
14-30ºC / 57-86ºF
20-80% rel. humidity, Non-Condensing
8000 ft (2438.4 Meters) max. altitude above sea level for use
Atmospheric pressure: 75-101 kPa
-30-60ºC / -22-140ºF
15-90% rel. humidity, Non-Condensing
No
≤75 decibels (dbA at 1meter)
40 - 150
100 - 2500 at start of procedure
100 - 9950 after procedure has started
Supplementary Information
English