Fluent Pro Fluid Management System
Verify that the Fluid Bag Clamp is open.
If bleeding is obscuring the visibility, the Intrauterine Pressure may need to be adjusted
during the procedure to provide a tamponade effect.
If performing a MyoSure procedure, advance the MyoSure Tissue Removal Device to the
fundus and allow circulation of fluid through the non-active blade to clear field.
Ensure the outflow stopcock is fully open.
Avoid over dilation of the cervix. If the cervix is over dilated, use a second tenaculum to seal
the cervix.
Verify the Fluid Bag Clamp is open.
Ensure the inflow tubing is not occluded or pinched.
Ensure the inflow stopcock is fully open.
Ensure the pressure setting is adequate.
Reprime the hysteroscope. Refer to Section 2.7.9.
Ensure the appropriate MyoSure TRD is being used for the tissue being resected.
Ensure the MyoSure TRD is flexible drive cable is fully inserted into the corresponding
connection on the MyoSure Panel of the Fluent Pro Fluid Management System.
Ensure the MyoSure TRD is fully inserted through the hysteroscope, and the entire cutting
window is visualized at the end of the hysteroscope.
Orient the MyoSure TRD cutting window against the tissue when the foot pedal is activated.
If you see flashing from the scope, it means the window is not pushing directly on the tissue.
Ensure the outflow tubing is not occluded or pinched.
Avoid applying excessive force to the MyoSure TRD handle.
Refer to scale calibration check, Section 6.1.1
WARNING!
If the deficit is rapidly increasing, or the visualization field does not respond to a change in
pressure set point, this may indicate that the uterus has been perforated or that fluid is
escaping elsewhere. Examine the visual field for injury or a leak from the cervix.
If the Fluent Pro Fluid Management System powers down or loses power unexpectedly, wait
at least 15 seconds then power on the system and follow on-screen prompts. To resume the
previous procedure, refer to Section 5.2.
If the system will not power on, verify the following:
WARNING!
Unplug the Power Cord from the wall outlet and from the system before checking the fuse.
Troubleshooting