6
Supplementary Information
Annual Inspection
6.1
The inspections detailed below are required to be performed annually to assess system functionality and
technical safety. Regular inspections will assist in early detection of possible malfunctions. This helps maintain
the system and increases its safety and service life. The inspections do not require a service technician but
must be performed by Biomedical Engineering.
Scale Calibration Check
6.1.1
The Scale Calibration Check is intended to verify that the Fluid Bag Scales and Waste Bag Scale meet
the specified tolerances. This test requires a precision weight (500g – 3000g).
6.1.1.1
6.1.1.2
6.1.1.3
72
Fluent Pro Fluid Management System
Press the Power Button on the top of the system to power on the system.
Touch the 'SETTINGS' icon.
Touch the scale calibration 'CHECK' icon. The Scale Calibration Check screen is displayed.
Figure 71: Scale Calibration Check Overview Screen
Supplementary Information
English