6.1.1.8
6.1.1.9
6.1.1.10 If the reading is not within the acceptable range, touch the back icon and proceed to re-zero
6.1.1.11 If the weight readings are still outside of the acceptable range, contact Technical Support.
6.1.1.12 If the weight reading is acceptable, remove the weight and repeat the step above for each
6.1.1.13 Touch the ' DONE' icon to return to the Settings screen, then select 'END' to return to the
74
Fluent Pro Fluid Management System
Once scales are zeroed, touch the 'NEXT' icon. The Scale Calibration Check Part 2: Check
Weight Readings screen is displayed.

Figure 74: Scale Calibration Check, Part 2 Screen

Hang a weight on one of the Hooks. The weight should read +/-5%.
the scales and repeat the weight reading steps.
scale.
Setup screen.