Resume Procedure (After Power Loss or Shutdown)
5.2
If the system is shutdown or a power loss occurs prior to completion of a procedure, the option of resuming the
last procedure will be available. Upon startup, the System Restarted screen is displayed. This screen displays last
procedure data, including Current Deficit, Fluid In and Cutting Time. Touch the 'RESUME PROCEDURE' icon to
resume the last procedure. Selecting 'NO' will start a new procedure and all previous procedure data will be lost.
If a new procedure is started unintentionally, the deficit will need to be calculated manually.
Troubleshooting
5.3
FloPaks not detected
5.3.1
5.3.1.1
Loss of Suction
5.3.2
5.3.2.1
5.3.2.2
5.3.2.3
Inadequate Suction
5.3.3
5.3.3.1
62
Fluent Pro Fluid Management System
Figure 69: System Restarted Screen
Firmly press towards the bottom of the FloPak to ensure the top of the FloPak is flush with
the front of the console and to engage the Lever.
Outflow
5.3.2.1.1
Ensure the hysteroscope outflow tubing luer lock on the Out-FloPak is securely
connected to the hysteroscope outflow channel.
5.3.2.1.2
Ensure nothing is obstructing or occluding the hysteroscope outflow tubing.
5.3.2.1.3
Reprime the hysteroscope. Refer to Section 2.7.9.
Under-Buttocks Drape
5.3.2.2.1
Ensure the Under-Buttocks Drape tubing suction connector is connected to the
Under-Buttocks Drape port and that the Under-Buttocks Drape is not occluding the
tubing.
5.3.2.2.2
Ensure nothing is obstructing or occluding the Under-Buttocks Drape tubing.
MyoSure TRD
5.3.2.3.1
Ensure the MyoSure TRD suction connector is securely connected to the
MyoSure TRD barb on the Out-FloPak.
5.3.2.3.2
Ensure nothing is obstructing or occluding the MyoSure TRD tubing.
Increase or decrease suction settings. Refer to Section 3.4 for instructions.
Troubleshooting
English