Item
Measurement range
Flow (mL/min)
Deficit (mL)
Accuracy
Pressure (mmHg)
Deficit (mL)
Dimensions (Width x Height x Depth)
System Weight
Console Mass
Safe Working Mass
Interfaces
Signal IN/OUT components
Mains connection
Maximum Load Cell Capacity
System Useful Life

Power Cord Safety

6.3
Ensure the connection data and technical specifications of the power supply comply with DIN VDE or national
requirements. The wall outlet power supply cord must be plugged into a properly installed safety wall plug (see
DIN VDE 0107). Read the device label located in rear of pump to determine the operating voltage of the system.
The power connection must be equipped with a grounding contact. Use the Fluent Pro Fluid Management
System power cord to establish a connection between the wall outlet and the power cord connection located in
the rear of the system.
Only for U.S. operators: Use only a certified (UL-listed), removable power cord, type SJT, minimal 16 AWG, 3
leads. The plug connectors must comply with NEMA 5-15 or IEC 320/CEE22. Grounding will only be reliable if
the equipment is connected to a corresponding hospital grade outlet.
Integrate the system into the potential equalization system as specified by local safety rules and regulations.
Medical devices are subject to special safety and protective measures concerning electromagnetic compatibility
(hereafter abbreviated as EMC).
This system is to be used only for the purposes described in the manual and must be installed, set up, and
operated in compliance with the EMC notes and instructions. See Section 6.4: Electromagnetic Compatibility.

Electromagnetic compatibility

6.4
Electromagnetic Emissions Guidance and Manufacturer's Declaration
6.4.1
The Fluent Pro Fluid Management System is intended for use in the electromagnetic environment
specified below. The customer or the user of the Fluent Pro Fluid Management System should ensure
that it is used in such an environment. Non-observance of the instructions listed in this manual can
lead to damage or malfunction of the system.
Use of accessories, transducers, and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
English
Fluent Pro Fluid Management System
Specification
0-1000
±9999
WARNING!
Supplementary Information
± 35
± 50 under normal use
23in x 67in x 27in / 584mm x 1702 mm x 686mm
65Kg
45Kg
20 Kg
None
IEC 60320-1 C14
<6.3Kg for Supply and Waste Bag Hooks
The system shall have a useful life up to 5 years.