Immunity test
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated HF interference
quantities according to
IEC 61000-4-3
Enclosure Port Immunity to
RF Wireless
Communications
English
Fluent Pro Fluid Management System
IEC 60601
Compliance level
test level
0% UT (100% dip
0% UT (100% dip
in the UT) for ½
in the UT) for ½
cycle at 0°, 45°,
cycle at 0°, 45°,
90°, 135°, 180°,
90°, 135°, 180°,
225°, 270°, and
225°, 270°, and
315°
315°
0% UT (100% dip
0% UT (100% dip
in the UT) for 1
in the UT) for 1
cycle and 70%
cycle and 70%
(30% dip in the UT)
(30% dip in the UT)
UT for 25/30 cycles
UT for 25/30 cycles
at 0°
at 0°
0% UT (100% dip
0% UT (100% dip
in the UT) for
in the UT) for
250/300 cycles
250/300 cycles
30 A/m
30 A/m
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
6 Vrms in ISM
Bands between
150kHz and
80 MHz
3 V/m
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
Per Table 9 of the
See table 7 for
IEC 60601-1-2
frequency ranges
standard
and test conditions
for RF wireless
communications
equipment
Supplementary Information
Electromagnetic environment
guidance
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user/operator of the Fluent Pro
Fluid Management System requires
continued operation during power mains
interruptions, it is recommended that the
Fluent Pro Fluid Management System be
powered from an uninterruptible power
supply or battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
3 Vrms
Portable and mobile RF communications
equipment used no closer to any part of
the Fluent Pro Fluid Management
6 Vrms in ISM
System,
Bands between
including cables, then the recommended
150kHz and
separation distance calculated from the
80 MHz
equation applicable to the frequency of
the transmitter.
Recommended safety distance:
d = 1.2√P for 150 KHz to 80 MHz d =
1.2√P for 80 MHz to 800 MHz d = 2.3√P
for 800 MHz to 2.7 GHz
Where P as the maximum output power
rating of the transmitter in watts [W]
according to the transmitter manufacturer
3 V/m
d as recommended separation distance
in meters [m].
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey a, should be
less than the compliance level in each
frequency range. b
Interference may occur in the vicinity of
equipment marked with the following
symbol: