During Procedure

Adjusting the Deficit Limit

3.1
3.1.1
About Deficit Limits
3.1.1.1
3.1.1.2
3.1.1.3
3.1.1.4
36
Fluent Pro Fluid Management System

Figure 44: Procedure Screen

The deficit is the total amount of fluid left in the patient or unaccounted for otherwise.
The default Deficit Limit at the start of a new procedure is set to 750 mL.
Before the procedure has started, the Deficit Limit can be adjusted from 100 mL to 2500 mL
in increments of 50 mL.
Once the procedure has started, the Deficit Limit can be increased beyond 2500 mL, if
necessary, at the discretion of the physician. The Fluid In is displayed on the procedure
screen in mL and reflects the total fluid in, including priming volume, unless the deficit was
zeroed out because system was not primed into the under buttocks drape.
During Procedure
English