Hang Fluid Bag
2.7.2
2.7.2.1
2.7.2.2
Unpack the Fluent Pro Procedure Kit
2.7.3
2.7.3.1
20
Fluent Pro Fluid Management System
Figure 17: Setup Step Detected Screen
Hang a Fluid Bag containing distension media appropriate for the procedure on the Fluid Bag
Hook. Hang only a single Fluid Bag on the Fluid Bag Hooks at a time. The Fluid Bag size
should not exceed 3L for each hook. If you need to replace a Fluid Bag during the
procedure, see Section 3.8.1.
Figure 18: Hang Fluid Bag
WARNING!
The Fluid Bag size should not exceed 3L per hook. Ensure the Fluid Bag is not resting on
top of console. Doing so may impact fluid deficit accuracy.
The Fluent Pro Fluid Management System does not need to be powered on to hang the
Fluid Bags. If the system is on, then follow the prompts. The system will detect and display a
'
DETECTED' icon on the Setup screen when the Fluid Bag has been properly installed.
Open the Fluent Pro Procedure Kit Packaging.
Procedure Setup
English