6.1.2.8
6.1.2.9
6.1.2.10 Verify the pressure is 26+/-5 in H2O.
6.1.2.11 If the value is not within the acceptable range, check for air in the tube and repeat the air
English
Fluent Pro Fluid Management System
Press 'STOP PURGE' once the tube is filled with fluid and there are no air bubbles
remaining.

Figure 79: Stop Purge for Pressure Calibration Check

Remove the end of the Inflow Tube from the container and hold it even with the tip of the
Fluid Bag bracket.

Figure 80: Performing Pressure Calibration Check

purge (Start Purge and Stop Purge). Verify the pressure is 26+/-5 inH2O. If the pressure is
still out of range, contact Technical Support.
Supplementary Information
77