GUI Notifications and Alerts
5.3.10
5.3.10.1 Notifications and Troubleshooting (Screen Flashing Yellow and Orange)

Table 1: Real-time Display, Notifications and Troubleshooting

Notification
Message
Fluid Bag(s) Low
Waste Bag Almost
Full
Approaching Deficit
Limit
High Fluid Loss
Reprime Suggested
MyoSure Device Not
Found
Excessive Torque
Detected
MyoSure Motor
Temperature High
MyoSure Motor
Overcurrent 1
MyoSure Motor
Overcurrent 2
MyoSure Motor
Under Voltage
MyoSure Motor
Stalled
English
Fluent Pro Fluid Management System
Description
Volume in Fluid Bag is less than 500mL
Volume in Waste Bag is greater than 5500mL
The deficit value is at or above 75% of the set
limit
The deficit value rose 300 mL or more over
the last 60 seconds of continuous run time
A change in pressure was detected due to a
hysteroscope change or incomplete prime
System does not detect that the MyoSure
device has been installed
MyoSure TRD speed is too slow
MyoSure TRD motor is too hot
MyoSure TRD current is too high
MyoSure TRD electrical issue
MyoSure TRD voltage is too low
MyoSure TRD motor stalled
Resolution
Add Fluid Bag with volume above 500mL.
Spike and unclamp the new bag.
Manually pause the procedure if running.
Remove full Waste Bag and replace with a
new empty Waste Bag. Resume procedure.
Consider increasing the deficit limit if clinically
appropriate or stopping the procedure.
Check for excessive cervical leakage. Verify
that the leakage is being captured in the
under-buttocks drape and returning to the
waste bag. Check for signs of uterine
perforation.
Reprime the system.
Ensure the MyoSure TRD is properly plugged
in and tap foot pedal to resume. If warning
persists, replace MyoSure TRD. Press clear
on the GUI.
Reduce pressure on device handle and tap
foot pedal to resume. If warning persists,
replace MyoSure TRD.
If warning persists, replace MyoSure TRD
If warning persists, replace MyoSure TRD
If warning persists, replace MyoSure TRD
If warning persists, replace MyoSure TRD
If warning persists, replace MyoSure TRD