Touch the 'START' icon. The Scale Calibration Check Part 1: Zero the Scales screen is
displayed.
Remove everything from the Fluid Hooks and Waste Bag Hooks.

Figure 72: Scale Calibration Check, Part 1 Screen

Touch the 'ZERO' icon.
When scales have been successfully zeroed, the scale values will read zero and a '
ZEROED' confirmation will be displayed. If the zeroing fails, follow prompts to retry. If zeroing
continues to fail, contact Technical Support.

Figure 73: Scale Calibration Check, Scales Zeroed Screen

Supplementary Information
SCALES
73