5.6. Insertion of Isiris α
Lubricate the insertion cord with a medical grade water based lubricant to ensure
the lowest possible friction when Isiris α is inserted into the patient's urethra. It is usual
practice to lubricate the patient's urethra using the same water based lubricant.
If the camera image of Isiris α becomes unclear the tip can be cleaned by removing
Isiris α and cleaning the tip with a piece of sterile gauze. Continue the procedure until
a satisfactory image is obtained.
5.7. Withdrawal of Isiris α
When withdrawing Isiris α, make sure that the deflection lever is in the neutral position.
Slowly withdraw Isiris α while watching the live image on Isiris Monitor.
By the end of procedure, put the distal tip in its neutral and non-angled position and
slowly withdraw the Isiris α without touching the bending lever. Always observe the
live endoscopic image while operating the grasper and manoeuvring and withdrawing
the Isiris α.
6. After use
Visual check
1.
Are there any missing parts on the deflective tip, lens, insertion cord, or grasper?
If yes, then take corrective action to locate the missing part.
2.
Is there any evidence of damage on the deflective tip, lens, insertion cord,
or grasper? If yes, then examine the integrity of the product and conclude
if there are any missing parts.
3.
Are there cuts, holes, sagging, swelling or other irregularities on the deflective
tip, lens, insertion cord, or grasper? If yes, then examine the product to conclude
if there are any missing parts. In case of corrective actions needed (step 1 to 3) act
according to local hospital procedures. The elements of the insertion cord and
grasper are radiopaque.
Final steps
Disconnect Isiris α from Isiris Monitor and dispose of Isiris α in accordance with local
guidelines for collection of infected medical devices with electronic components.
Switch off Isiris Monitor by pressing the power button for at least 2 seconds.
7. Trouble shooting
If problems occur with the Isiris α, please use this trouble-shooting guide to identify the
cause and correct the error.
Absent or reduced aspiration capability.
Cause
Irrigation channel blocked
492880500_Isiris IFU_Multi_V01_202205_TCC 11251.indb 17
492880500_Isiris IFU_Multi_V01_202205_TCC 11251.indb 17
Action
Clean the irrigation channel by flushing the irrigation
channel with sterile saline using a syringe.
17
02/06/2022 17:47
02/06/2022 17:47