The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
775 warming unit should assure that it is used in such an environment.
Emissions test
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
Flicker emissions
IEC 61000-3-3
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the
Model 775 warming unit should assure that it is used in such an environment.
Immunity Test
Electrostatic
±8 k V contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 k V for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line to line
IEC 61000-4-5
±2 k V line(s) to earth
Voltage dips, short
<5% UT
interruptions and
(>95% dip in UT)
voltage variations on
for 0,5 c ycle
power supply input
lines
40% UT
IEC 61000-4-11
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
10
Guidance and manufacturer's declaration – electromagnetic emissions
Compliance
The Model 775 warming unit uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
The Model 775 warming unit is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Guidance and manufacturer's declaration – electr omagnetic immunity
IEC 60601
Compliance
test level
±8 k V contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line to line
±2 k V line(s) to earth
<5% UT
(>95% dip in UT)
for 0,5 c ycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
Electromagnetic environment - guidance
Electromagnetic environment - guidance
level
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that o f a typical commercial or
hospital environment.
Mains power quality should be that o f a typical commercial or
hospital environment.
Mains power quality should be that o f a typical commercial or
hospital environment. I f the user of the M odel 775 warming
unit requires continued operation during power mains
interruptions, it is recommended that the M odel 775 warming
unit be po wered from an unin terruptible power supply or a
battery.
Power frequency magnetic fields should be at le vels
characteristic of a typical location in a typical commercial or
hospital environment.