SLOPE
,
,
Type:
Setting or Query
Setting Syntax:
SLO NEG
POS
Examples:
SLO POSITIVE
SLOPE PCIS
SLOPE NEGATIVE
SLO NEG
Query Syntax:
SLO?
Query Response:
SLO POS;
i -
SLO NEG;
Discussion:
The SLOPE command sets the input trigger for the se-
lected channel to the specified slope. The valid arguments
are:
NEG
-
lnput will trigger on negative going edge.
POS
-
lnput will trigger on positive going edge.
The power-on initial setting is SLO POS.
For information on selecting channels see discussion of
the CHANNEL. command.
SLOPE
Operating Instructions-DC 50 10
START
Type:
Operational
Syntax:
START
Discussion:
The START command starts a TMANUAL or TOTALIZE
A, TOTALIZE A+B, TOTALIZE A-B, measurement. For
EVENTS, FALL, FREQUENCY, PERIOD, RATIO, RISE,
TIME, or WIDTH measurements, START restarts measure-
ment if STOPped.
START