Operating Instructions-DC
501 0
ERROR
-
Type:
Query
Syntax:
ERR?
ERROR?
Response:
ERR <number>;
Discussion:
The ERRlOR query is used to obtain information about
the status of the instrument.
If RQS is ON, the ERROR query returns an event code
<number> describing why the RQS bit was set in the last
Status Byte reported by the instrument. The event code is
then reset to
0.
If RQS is OFF, the ERROR query returns an event code
,
-
<number>
describing the highest priority condition cur-
rently pending in the instrument. This event code is then
cleared and another ERROR query will return the event
code for the next highest priority condition pending.
ERROR
EVENTS (EVENTS B DURING A)
Type:
Operational
Syntax:
EVE BA (argument is optional)
Examples:
EVENTS BA
EVE
Discussion:
The EVENTS command sets up the DC 501 0 to measure
the total number of events occurring on Channel
B
during
the pulse width of the input signal on Channel A.
EVENTS (EVENTS B DURING A)