Calibration-DC
SO
10
Performance Check Procedure
12. Check lnput Sensitivity:
X I Attenuation, DC
and AC coupled; 1 MJZ, <70 mV p-p at G200 MHz
-
Refer to Fig. 4-5, performance check setup. Use the fol-
lowing control settings.
Sinewave Generator
Frequency Range
100-250
Frequency Variable
200
Amplitude Multiplier
X 1
Output Amplitude
0.70
DC 5010
RATIO B/A
(lighted)
a. Connect the 1 MHz frequency standard to the
DC 501 0 CHANNEL A input.
b. Connect a coaxial cable with a 1 0 x attenuation and
50
Q
termination from the sinewave generator Output to the
DC 501 0 CHANNEL B input.
13. Check lnput Sensitivity: X5 Attenuation, DC
and AC Coupled:
1
MQ, ~ 5 9 4 mV p-p at ~ 3 0 0 MHz
Refer to Fig. 4-5 performance check setup. Use the fol-
lowing control settings.
Sinewave Generator
Frequency MHz
300
Amplitude
&94
mV p-p
DC 5010
CHANNEL A and B
TERM
1 MQ (unlighted)
ATTEN
X5 (unlighted)
a. Remove the DC 5010 CHANNEL B input connection
and insert a 50
Q
termination to the coaxial cable. Recon-
nect this cable to the CHANNEL B input.
b. Press the DC 501 0 AUTO TRlG button, LEVEL CH B
button, and the DISPLAY-TEST button.
c. Press the DC 5010 AUTO TRlG button, LEVEL CH B
button, and the DISPLAY-TEST button.
c. Press the DC 501 0
f
(increment) or
1
(decrement) but-
ton to adjust the trigger level for a stable display readout.
d. Press the DC 501 0 (increment) or
1
(decrement) but-
ton to adjust the trigger level for a stable display readout.
e. CHECK-that
the DC 5010 display readout indicates
approximately 200.00XXX (the last three digits can vary due
to source instability).
f. Press to light the DC 5010 FREQ A button.
d. CHECK-that
the DC 5010 display readout indicates
approximately 300.00XXX (the last three digits can vary due
to source instability).
e. Remove the 1 MHz frequency standard from the
DC 5010 CHANNEL A input.
f. Move the DC 5010 CHANNEL B input connection to
g. Remove the 1 MHz frequency standard from the
the CHANNEL A input.
DC 501 0 CHANNEL A input.
h. Move the DC 5010 CHANNEL B input connection to
g. Press to light the DC 501 0 FREQ A button, AUTO
the CHANNEL A input.
TRlG button, LEVEL CH A button, and the DISPLAY-TEST
button.
i. Press the DC 501 0 AUTO TRlG button, LEVEL CH A
button, and the DISPLAY-TEST button.
h. Press the DC 501 0
f
(increment) or
1
(decrement) but-
ton to adjust the trigger level for a stable display readout.
j. Press the DC 5010
f
(increment) or
1
(decrement) but-
ton to adjust the trigger level for a stable display readout.
i. CHECK-that
the DC 501 0 display readout indicates
approximately 300.00XXXX (the last four digits can vary
due to source instability).
k. CHECK-that
the DC 501 0 display readout indicates
approximately 200.00XXXX (the last four digits can vary
due to source instability).
j. Remove the cable connections.
REV JUL
1983