SO 10
Refer to the power module instruction manual and make
sure that the line jumpers are positioned correctly for the
line voltage in use. Check the counter and the power module
for the proper fuses. Be certain that the power plug for the
power module has the proper grounding conductor.
C A U T I O N
a
To prevent damage to the instrument, turn the power
module off before installation or removal from the
power module. Do not use excessive force to install or
remove the instrument from the power module.
Check to see that the plastic barrier keys on the
interconnecting jack of the selected powei. module compart-
ment match the cutouts in the rear interface connector for
the counter. If they do not match, do not insert the counter
until the reason is investigated.
If the cutouts and barrier keys match, align the chassis of
the counter with the upper and lower guides of the selected
--
compartment. See Fig. 2-1
.
Insert the counter into the com-
partment and press firmly to seat the rear inteirface connec-
tor. Apply power by operating the POWER switch on the
power module.
The plastic lockouts (see Fig. 2-1) prevent programmable
instruments from being used in the TM 500-Series (manual
instruments) Power Module.
To remove the counter from the power module, turn off
the POWER switch, pull the release latch knob (located in
the lower left front corner) until the interconnecting jack dis-
engages. Pull the counter straight out of the power module
compartment.
FRONT PANEL OPERATION
The following information is a brief, functional description
FRONT PANEL DISPLAY
of the front panel display, controls, and connectors (See
Fig. 2-2).
@
Display
The display contains nine seven-segmenls LEDS and
eight annunciators. All measurement results (are displayed
with the best possible resolution. The readout (result) for the