Herunterladen Inhalt Inhalt Diese Seite drucken

Tektronix DC 5010 Anleitung Seite 56

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operating Instructions-DC
SO 10
IDENTIFY
Type:
Query
Syntax:
ID?
IDENTIFY?
Response:
ID TEK/DCSOlO,V79.l ,Fx.y ;
Discussion:
The IDENTIFY query returns the above response where:
TEKIDC 501 0
-
Identifies the instrument type.
V79.1
-
Identifies the version of Tektronix
Codes and Format Standard to which
the instrument conforms.
-
ldentifies the firmware version of the
instrument, where x.y is a decimal
number.
IDENTIFY
INITIALIZE
Type:
Operational
Syntax:
IN IT
INITIALIZE
Discussion:
The INIT command performs a power-on ir~itialization of
the instrument's settings. The power-on settings for the
DC 501 0 are:
FREQ A
AVE -1
FIL OFF
NULL OFF
SLO POS (Channels A&B)
ATT 1 (Channels A&B)
COU DC (Channels A&B)
TERM HI (Channels A&B)
CHA A
OPC OFF
OVER OFF
PRE OFF
DT OFF
USER OFF
RQS ON
In addition, an autotrigger is performed to set the trigger
levels. With the maximum and minimum peak values of
autotrigger performed, the maximum execution time for the
INlT function is 2.5 seconds.
The INIT command does not generate a power-on SRQ
nor does it put the instrument in LOCAL mode! as a normal
power-on does.
INITIALIZE

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis