PROBECOMP (PROBE COMPENSATION)
-
Type:
Operational
Syntax:
PROBE I~&B (argument is optional)
Examples:
PROBECOMP A&B
PROB
Discussion:
The PROIBE COMP command sets up the DC 5010 to
provide information which can be used to help compensate
probes.
This function generates 2-digit results. The most signifi-
cant digit is the result for Channel A and the least significant
digit is the result for Channel B.
The PROBECOMP function uses the autotrigger oper-
,
ation in the compensation process. Therefore, trigger levels
and MIN and MAX values will be affected by PROBECOMP
measurements.
The autotrigger, used by PROBECOMP, is a fast version
of auto, with fmi, approximately 100 Hz and maximum ex-
ecution time approximately 0.25 second. This fast auto may
be used to quickly update MIN and MAX values for signals
greater than 100 Hz.
For more information see description of Probe Compen-
sation in this manual.
PROBECOMP (PROBE COMPENSATION)
Operating Instructions-DC 501 0
RATIO
Type:
Operational
Syntax:
RAT B/A
Examples:
RATIO B/A
RAT
Discussion:
The RATIO command sets up the DC 5010 to measure
the ratio of events on Channel B to the events on
Channel A.
RATIO