Calibration-DC
501 0
Performance Check Procedure
c. CHECK-that
the
DC5010
readout is within
b. Press the DC 5010 AUTO TRIG button, and then
999.99879 kHz and 1.0000020 MHz.
press LEVEL CH A.
L '
d. Remove all cable connections from the DC 501 0.
c. Press the DC 5010 (increment) button until the digi-
tal multimeter display just changes to a low readout, if it is
not already low. Press the DC 5010
4
(decrement) button
until the digital multimeter display just changes to a high
readout.
3. Check the Trigger Level CH A and CH B
Accuracy
d. CHECK-that
the DC 501 0 display readout indicates
Refer to Fig. 4-1, performance check setup. Use the fol-
b e h ~ e e ~
+0-040 and -0.040.
lowing control settings.
Digital
Multimeter
Function-Range
2 V
DC
5010
CHANNEL A and
CHANNEL B
ATTEN
X I (lighted)
e. Connect a coaxial cable from the pulse generator out-
put to the digital multimeter input using a bnc-to-banana
adapter.
f. Set the pulse generator for Ext Trig and Ext Dur and
adjust the output for a displayed readout of approximately
+
1.97 on the digital multimeter. Note this reading.
a. Connect the DC 501 0 CH A SHAPED OUT to the digi-
g. Disconnect the cable from the digital multimeter bnc-
tal multimeter input using the bnc-to-slide on connector as-
to-banana adapter and connect to the DC 5010 CHANNEL
sembly (standard accessory).
A input.
Power
Module
\
Pulse
Generator
Digital
Multimeter
Fig. 4-1. Performance Check setup for step 3.
REV JUL
1983